 Air Filter
Systems marked by the Service Revisions listed in the table below are equipped with
new air filters. The Filters are intended to reduce dust and debris entering and
accumulating inside the System. Dust and debris may affect internal heat exchanges
and the performance of other components.
Platform Description
Part Number
Service Revision
InMode II
AG606666A
B04
Optimas
AG606991A
B04
Triton
AG606992A
A02
Dust filter are located on inside back over door (Figure -13).
Figure -13: Back Cover Door with Dust Filter
Air Filter Maintenance
Dust Filter must be to be cleaned once a week. In dusty environments or desert
regions, it should be checked daily.
Replace the Dust Filter if damaged.
Cleaning method
Remove back cover door and wash it under tap water until filter net is clean from
a dust. Air pressure can used carefully if available.
Shake and dry the back-cover door before use.
Return the back-cover door when dry.