16
Qubit
4 Fluorometer User Guide
Guidelines for using the Qubit
4 Fluorometer
Recommendations
To obtain the best results, follow the recommendations below. For more
information, see “Critical Qubit
Assay considerations, page 57.
Do not operate the instrument in direct sunlight.
Wear gloves during sample handling.
Use the instrument at room temperature only (2228ºC).
Bring all kit reagents to room temperature and insert all assay tubes into the
instrument only for as much time as it takes for the instrument to measure the
fluorescence.
Do not hold the assay tubes in your hand before performing a measurement.
Make sure that you have calibrated the Qubit
4 Fluorometer using the
appropriate standards.
Incubate the tubes for the Qubit
DNA and RNA assays for 2 minutes after
mixing the sample or standard with the working solution.
Incubate the tubes for the Qubit
protein assays for 15 minutes after mixing
the sample or standard with the working solution.
If you are performing multiple readings of a single tube, remove the tube
from the instrument and let it equilibrate to room temperature for 30 seconds
before taking another reading.
Note: Multiple readings of RNA samples is not recommended.
Visit thermofisher.com/qubit
for additional application notes, technical notes,
citations, software updates, and a list of validated Qubit
assays that have
been tested using the Qubit
4 Fluorometer.
Assay tubes for the
Qubit
4
Fluorometer
Only thin-wall, clear 0.5-mL PCR tubes are appropriate for use in the Qubit
4
Fluorometer. Acceptable tubes include Qubit
assay tubes (Cat. No. Q32856,
500 tubes). The minimum assay volume must be 200 µL for an accurate read.