EasyManua.ls Logo

Invitrogen Qubit 4 - Troubleshooting; Sample Handling and Reading Issues

Invitrogen Qubit 4
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Qubit
4 Fluorometer User Guide
55
Appendix A: Troubleshooting
Troubleshooting
Handling samples
The calibration standards included in the Qubit
microRNA, Qubit
RNA HS,
Qubit
RNA XR, Qubit
RNA IQ, and Qubit
RNA BR Assay Kits are high-
quality RNA standards. The integrity and concentration of these standards is
critical to the optimal performance of the Qubit
RNA assays. We highly
recommend treating the rRNA standards as you would any other precious
RNA. Use appropriate RNAse-free handling techniques, including RNAse-
free gloves, pipette tips, and tubes. Keep the tube lids closed whenever
possible; do not press the pipet to the inside wall of the tube when
withdrawing a sample. Return the RNA standards to 80°C as soon as
possible after use.
Ensure that the assay tubes are at room temperature at the time the reading is
taken. Do not hold assay tubes in your hand and do not leave assay tubes in
the Qubit
4 Fluorometer for longer than it takes to read the fluorescence. See
Assay temperature, page 58.
Be careful not to spill sample into the sample chamber. Promptly wipe any
spills.
The Qubit
assays are very sensitive and even small amounts of material
from a previous sample may result in errors. Use a clean 0.5-mL PCR tube for
each reading.
The tube must be clean and dry on the outside when taking readings.
Moisture and condensation on the tube surface can lead to reading errors.
Minute bubbles in samples will cause errors in readings. Be sure not to
introduce bubbles into samples. Slight tapping on the tube wall or brief
centrifugation will often help dissipate bubbles.
High reading
The sample is out of range. Use a sample that is less concentrated.
For Qubit
quantitation assays, view the
Fluorescence vs. Concentration graph in the
Results screen to confirm that the values for
the samples fall between the values of the
standards (page 27).
Ensure that the lid is closed while reading
standards and samples.
Prepare samples and standards according to
the instructions in the Qubit
assay kit you
are using.
Ensure that the assay is performed entirely at
room temperature.

Related product manuals