Read samples
Before you begin
Calibrate the Qubit
4 Fluorometer as described on page 18. (Run the
appropriate standards or accept the values from the previous calibration.)
Prepare the samples. Refer to the instructions provided with the assay.
Note: Incubate the samples for the appropriate amount of time after
mixing them with the working solution (2 minutes for the Qubit
DNA
and RNA assays, 15 minutes for the Qubit
protein assay).
Read samples
1. In the Read standard (for Qubit
quantitation assays) or Ready for samples
screen (for Qubit
RNA IQ assay), press Run samples.
2. In the Sample volume screen, select the sample volume and units for the Qubit
quantitation assays or the Qubit
RNA IQ assay:
a. Press the + or buttons on the wheel to select the sample volume added
to the assay tube (between 1 and 20 µL).
b. For the Qubit
quantitation assays, select the units for the output sample
concentration from the dropdown menu.
Note: In the RNA IQ assay, there are no units associated with the RNA IQ
value (i.e., RNA IQ#) and only sample volume is required.