Qubit
4 Fluorometer User Guide
Device verification test
Perform Device
verification test
The device verification test checks the internal components of the Qubit
4
Fluorometer. Perform the device verification test when a problem with the
instrument is suspected. It is not necessary to perform the verification test on a
regular basis.
1. On the Home screen, press Settings.
2. On the Settings screen, press Device verification test. Verification test screen
opens.
3. On the Verification test screen, press Start test.
4. When the test is complete, the software displays the error status.
If no errors are found, press Close to return to the Settings screen.
If errors are found, a Verification test failed message appears. Verify that
the test was run with the lid closed. If the Verification test failed message
persists with the lid closed, contact Technical Support for help (page 69).