Calibrate assays on the Qubit
4 Fluorometer
Introduction
For each assay, you have the choice to run new standards for calibrating the assay
on the Qubit
4 Fluorometer or to use the values from the previous calibration (for
more information, see “
Qubit™ 4 Fluorometer, page 58). This section provides
instructions for running new calibration standards and for using the last
calibration.
Read new
standards for
calibration
1. On the Home screen, select the assay type for which you wish to read new
standards.
To go to the second page of available assays, swipe the screen to the right. To
return to the first page, swipe the screen to the left.
2. Select the desired assay.
Note: The software displays the available assays for the assay type
you selected in step 1. The example in step 2 shows the available
assays for the dsDNA assay type.