56
Qubit
4 Fluorometer User Guide
Low reading
The sample is out of range. Use a sample that is more concentrated or use a
lower dilution (for example, 20 µL in 180 µL instead of 10 µL in 190 µL).
For Qubit
quantitation assays, view the
Fluorescence vs. Concentration graph in the
Results screen to confirm that the values for
the samples fall between the values of the
standards (page 27).
Ensure that you have prepared the Qubit
working solution correctly (1:200 dilution
using the buffer provided in the kit).
Ensure that you have prepared the standard
tubes correctly (10 µL of each standard in
190 µL of Qubit
working solution).
Ensure that the standard and sample tubes
are filled to 200 µL.
Protect the Qubit
reagent and working
solutions from light.
Select the correct Qubit
4 Fluorometer assay for the Qubit
assay you are
performing and calibrate the fluorometer correctly. Standards must be used
in the correct order.
Ensure that the assay is performed entirely at room temperature.