IONTECH
®
IT-DS EDI Modules
Screwdrivers (flat blade and Phillips head)
IT-DS module handles for moving & positioning unit (IONTECH part number
IT-DSHANDLES)
2.2 Electrical and Plumbing Supplies
The amounts, sizes, and types of these supplies will vary from site to site. Check before hand to
determine the site needs.
Conduit, wires and appropriate conduit connectors to run the DC power and ground from
the DC power controller to the module. Size wire in accordance with local electrical code.
Fittings to connect the 1 1/4" and 3/4" BSPM ports on the module to the module
Grounding Module Piping - To avoid the risk of electrical shock, some
form of grounding must be used on any stream where the plumbing is
stainless steel or if there are samples points or instrumentation in close
proximity to the module. For sanitary applications, a grounding cap can be
used, IONTECH part number 49/S4501-124, which is actually a 3/4"
TC cap with a welded stud to be wired to ground. For non-sanitary
applications, a 1/4" SS threaded grounding rod can be used, IONTECH
part number ZIVC00160 see Figure 2-1.
Do not uncrate the module prior to moving it into its final location. After uncrating it, inspect it for
any possible signs of damage. If damage is apparent, immediately notify your Local Service
Provider and the carrier.
2.4 Operating Requirements
In order to operate to specification, IT-DS modules must have the following conditions present. If any
of these conditions are unmet, do not attempt to install IT-DS modules without specific instructions
from your Local Service Provider Technical Support.
2.4.1 Operating Environment
IT-DS modules require indoor installation out of direct sunlight. The maximum ambient room
temperature should not exceed 113 °F (45° C). The module can tolerate humidity of up to 90%,
as long as condensation does not occur.