EasyManua.ls Logo

ION TIGER XT - Technical Specifications

ION TIGER XT
71 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TIGER XT Instrument User Manual V1.2
Pioneering Gas Sensing Technology. Page 68 of 71 ionscience.com
11. Technical Specifications
Minimum Resolution
10.6 eV: 0.001 ppm (1 ppb)***
11.7 eV: 0.6 ppm (600 ppb)***
Detectable Range**
10.6 eV: 1 ppb up to 20,000 ppm***
11.7 eV: 9,000 ppm***#
Response Time
10.6 eV: T90 < 2 seconds
11.7 eV: < 6 seconds
Accuracy
10.6 eV: +/- 5% display reading or +/- one-digit ***
11.7 eV: +/- 12% display reading ***
Intrinsically Safe Approvals
II 1G Ex ia IIC T4 Ga
Tamb = -25
o
C ≤Ta ≤+45
o
C (with Lithium-ion Battery Pack)
Tamb = -25
o
C ≤Ta ≤+40
o
C (with Alkaline Battery Pack)
IECEx ITS 22.0025X ITS-I22ATEX35111X
ITS22UKEX0635X
3193491 conforms to UL Std. 913, 61010-1 &
Certified to CAN/CSA Std. C22.2 No. 61010-1
Battery Life (Li-ion)
Typical operating time of 24 hours
Charge time typically 8 hours
Lamps
10 eV Krypton
10.6 eV Krypton (standard)
11.7 eV argon
Data Logging
>120,000 data log points including date and time stamp
Communication
Direct USB 1.1
Calibration
2-point and 3-point calibration (via calibration kit accessory)
Alarms
Flashing LEDs’s amber (low alarm) red (high alarm)
Sounder 95 dBA at 300 mm (12”)
Vibration on alarm
Pre-programmed TWA and STEL
Flow Rate
220 ml/min
Humidity
0-99% RH (non-condensing)
Protection
Designed to IP65
EMC tested to EN61326-1:2013 & EN50270:2015 & CFR 47:2008
Class A

Table of Contents