11. Technical Specifications
10.6 eV: 0.001 ppm (1 ppb)***
11.7 eV: 0.6 ppm (600 ppb)***
10.6 eV: 1 ppb up to 20,000 ppm***
11.7 eV: 9,000 ppm***#
10.6 eV: T90 < 2 seconds
11.7 eV: < 6 seconds
10.6 eV: +/- 5% display reading or +/- one-digit ***
11.7 eV: +/- 12% display reading ***
Intrinsically Safe Approvals
• II 1G Ex ia IIC T4 Ga
• Tamb = -25
o
C ≤Ta ≤+45
o
C (with Lithium-ion Battery Pack)
• Tamb = -25
o
C ≤Ta ≤+40
o
C (with Alkaline Battery Pack)
• IECEx ITS 22.0025X ITS-I22ATEX35111X
• ITS22UKEX0635X
• 3193491 conforms to UL Std. 913, 61010-1 &
Certified to CAN/CSA Std. C22.2 No. 61010-1
Typical operating time of 24 hours
Charge time typically 8 hours
10 eV Krypton
10.6 eV Krypton (standard)
11.7 eV argon
>120,000 data log points including date and time stamp
2-point and 3-point calibration (via calibration kit accessory)
Flashing LEDs’s amber (low alarm) red (high alarm)
Sounder 95 dBA at 300 mm (12”)
Vibration on alarm
Pre-programmed TWA and STEL
0-99% RH (non-condensing)
Designed to IP65
EMC tested to EN61326-1:2013 & EN50270:2015 & CFR 47:2008
Class A