This step allows you to copy an existing meter setup and modify it to create the new Intelis
Gas Meter (for example, choose I-250, Itron). Select a similar meter's row. Follow the Meter-
Table instructions: Click on the next open/empty row. To complete the addition, click off of
the table and select Add Chosen Meter. The following illustration shows the successful copy
of the I-250 meter to create the Intelis 250 MeterName, ITRON-RTY MeterMfr.
8. Update the parameters in the table.
MeterName: Intelis 250
MeterMfg: ITRON-RTY
Open proof: 100
Check proof: 100
Other proof: 100
Open flow: 250 (or the setting required by your utility)
Check flow: 50 (or the setting required by your utility)
Other flow: 100
Open revs: 6
08 July 2020 815-0154-00 REV 000 Itron, Inc. Page 24 of 66
Intelis Gas Meter Proving and Setup Guide
4 Proving the Intelis Gas Meter Using the Honeywell SNAP Proving