36
Index
Category
Error ERR Conti- Error Message Text
No. nuation
possible
60 System The vacuum does not 802 No vacuum increase could be measured. Check the following points: Is the firing chamber tight (no
increase (self-test) contamination on the sealing surfaces)? Is the vacuum hose connected? Is the vacuum pump connected?
Is the fuse F1 o.k.?
61 System Temp. EXTERNAL T-SENSOR 1010 Temperature channel EXTERNAL T-sensor excess temperature
excess temperature
(> 1225°C)
62 System Write error in the furnace 1011 Error while saving furnace configuration data to the internal memory.
configuration data memory
63 System Read error in the furnace 1012 Error while reading the furnace configuration data from the internal memory.
configuration data memory
64 System Checksum error in the 1013 Invalid checksum of the memory - furnace configuration data are written into the internal memory using
furnace configuration the original values.
data memory
65 System Write error in the furnace 1014 Error while saving the furnace operational data to the internal memory.
operational data memory
66 System Read error in the furnace 1015 Error while reading the furnace operational data from the internal memory.
operational data memory
67 System Checksum error in the 1016 Invalid checksum of the memory – furnace operational data are written into the internal memory using
furnace operational data the original values.
memory
68 System Write error in the firing 1017 Error while saving the firing program protocol data.
protocol data memory
69 System Read error in the firing 1018 Error while reading the firing program protocol data.
protocol data memory
70 System Checksum error in the 1019 Invalid checksum of the memory – firing program protocol data are deleted.
firing protocol data memory
71 System Technical error 1024 no Error while reading the stop switch for the furnace head.
of the furnace head **,***
72 System Technical error of the 1025 no Read/write CPLD
furnace head
**, ***
73 System Technical error of the 1026 no Error in the vacuum driver
vacuum driver
**, ***
74 System Technical error in the 1028 Error while reading/writing the SRAM.
SBS driver
75 System Write error in the firing 1143 Error while saving firing program data to the USB memory stick.
program memory
76 System Read error in the firing 1144 Error while reading firing program data from the USB memory stick.
program memory
77 System Checksum error in the 1145 Invalid checksum of the memory for firing program data – the firing program data are written to the USB
firing program memory memory stick using original values.
78 System Write error in the firing 1146 Error while saving firing group data to the USB memory stick.
group memory
79 System Read error in the firing 1147 Error while reading the firing group data from the USB memory stick.
group memory
80 System Checksum error in the 1148 Invalid checksum in the memory for firing group data - the firing group data are written to the USB
firing group memory memory stick using original values.
81 System Reading, calculating the 1202 no Error while measuring the ambient temperature
ambient temperature
**,***
82 System Reading, calculating the 1203 no Error while measuring the furnace temperature
furnace temperature
**,***
83 System Reading, calculating the 1204 no Error while measuring the furnace control temperature
furnace control temperature **,***
84 System Reading, caluclating the 1205 Error while measuring the resistance value for the ATK2 calibration.
resistance value
85 System Reading, calculating, 1206 Error while measuring the temperature for the EXTERNAL T-sensor.
EXTERNAL T-sensor
86 System Temperature regulator 1207 no Error in the temperature regulator.
**,***
87 System ATK2 calibration: 1300 Error during calibration.
Pre-heating to 660 °C**
88 System ATK2 calibration: 1301 Error during calibration.
Calibration of 660 °C **
89 System ATK2 calibration: 1302 Error during calibration. Sample may not be correctly inserted. Try again with a new sample and make sure the
Pre-heating to 963 °C ** sample makes ample contact.
90 System ATK2 calibration: 1303 Error during calibration.
Calibration of 963 °C **
91 System
ATK2 calibration: Difference
1304 Error during calibration.
in the calibration values **
92 System ATK2 calibration: 1305 Error during calibration.
Calibration value range **
93 Note Calibration reminder 1310 Some time has passed since the last calibration procedure. Calibrate the furance soon.