37
Index
Category
Error ERR Conti- Error Message Text
No. nuation
possible
95 Note Dehumidification
reminder
1312 Some time has passed since the last dehumidification. Conduct a dehumidification in the near future.
96 System Access Board Descriptor: 1400 Error while writing the new version number to the BoardDescriptor (E2Prom).
Writing the version number
97 System Access Board Descriptor: 1401 Error while writing the new serial number to the BoardDescriptor (E2Prom).
Writing the serial number
98 System Access Board Descriptor: 1402 Error while reading the serial number from the BoardDescriptor (E2Prom).
Reading the serial number
99 System
Loading of the drivers failed
1500 no Failure during loading the necessary drivers. The furnace is not ready. Contact your Ivoclar Vivadent Service
*** Center.
100 Note Temperature > VT at the 1510 The temperature in the firing chamber is higher than the pre-drying temperature. Press START to continue
start of a firing program the program despite the error message.
101 Entry Software update: 1520 The file selected for the software update is invalid. The update cannot be conducted.
Invalid file
102 Entry Software update: 1521 The software to be installed is older or the same as the one installed on the furnace. The update cannot be
Invalid version conducted.
103 System Software update: 1522 An error has occurred during the software update. Do NOT switch off the furnace and try again.
Error during update If the error reoccurs, try conducting the update via the USB interface.
104 System LED-Print: Error during 1530 An error has occurred upon communication with the LED printed board.
communication with the
LED printed board
106 Inform. Change operation mode 1550 The operation mode has been changed! Make sure that the furnace runs at the new stand-by temperature
long enough before you start a program.
Behaviour of the furnace in case of error
* Furnace head opens when this error occurs.
** A program in progress is stopped.
*** The error cannot be acknowledged; the programs cannot be started.