SPECIFICATIONS
Model 3000 Operation Manual 33
Pacer Pulse Rejection
Width: 0.1 to 2 ms at ±2 to ±700 mV
Overshoot: Between 4 to 100ms and not greater than 2mV.
Fast ECG signals: 2mV/100µs.
Detector disabling: None.
NOTE: Pacemaker pulses are not present in any rear panel outputs.
Alarms
High Rate: 15 to 250 bpm in 5 bpm increments
Low Rate: 10 to 245 bpm in 5 bpm increments
Asystole: R to R interval >6 seconds
Lead Off: Detached lead or offset potential >0.5 V
Test Mode
ECG: 1 mV/100 ms @ 70 bpm
Display
Type: Active Matrix TFT Color LCD (640x480)
Trace: Single ECG trace with “freeze” function.
Screen Size: 13.25cm x 9.94cm, 16.5cm (6.5in)diagonal
Sweep Speed: 25, 50 mm/s
Aspect ratio: 0.4 (standard). User selectable.
Mechanical
Size: Height: 6.70in. (17.2cm)
Width: 9.25in. (33.5cm)
Depth: 9.21in. (23.4cm)
Weight: 6.5lbs (2.9kg)
Recorder (optional)
Writing Method: Direct Thermal
Number of Traces: 1
Modes: Direct - Manual Recording
Timed - Print button initiates a 30 second recording
Delay - Recorder is activated by an alarm
Paper Speeds: 25 and 50 mm/s
Resolution: Vertical - 200 dots/in.
Horizontal - 600 dots/in. at 25 mm/s
400 dots/in. at >25 mm/s
Frequency Response: >100 Hz at 50 mm/s
Data Rate: 400 samples/s/trace

Other manuals for Ivy Biomedical Systems 3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ivy Biomedical Systems 3000 Series and is the answer not in the manual?

Ivy Biomedical Systems 3000 Series Specifications

General IconGeneral
BrandIvy Biomedical Systems
Model3000 Series
CategoryMonitor
LanguageEnglish

Related product manuals