SAFETY
Model 7600/7800 Operation Manual 9
Guidance and manufacturer’s declaration Electromagnetic immunity
The Model 7600/7800 monitor is intended for use in the electromagnetic environment specified
below. The customer or the user of the Model 7600/7800 should ensure that it is used in such an
environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8kV air
±9 kV contact
±12kV air
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
Transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output lines
±3 kV for power
supply lines
±1.5 kV for
input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1.5 kV
differential mode
±3 kV common
mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines
IEC61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
) for
5 cycles
70 % U
T
(30 % dip in U
T
) for
25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 sec cycle
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 sec cycle
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Model 7600/7800 requires
continued operation during power
mains interruptions, it is
recommended that the Model
7600/7800 be powered from an
uninterruptible power supply.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
10 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ivy Biomedical Systems 7600 and is the answer not in the manual?

Ivy Biomedical Systems 7600 Specifications

General IconGeneral
BrandIvy Biomedical Systems
Model7600
CategoryMedical Equipment
LanguageEnglish

Related product manuals