SAFETY
Model 7600/7800 Operation Manual 5
5.0 SAFETY
5.1 Electrical
This product is intended to be operated from a mains power source of 100-120V~ or 200-230V~, 50/60 Hz and a
maximum ac power consumption of 45VA.
WARNING: To avoid risk of electric shock, this equipment must only be connected to a supply mains with
protective earth. Connect the monitor only to a three-wire, grounded, hospital grade receptacle. The three-conductor
plug must be inserted into a properly wired three-wire receptacle; if a three-wire receptacle is not available, a
qualified electrician must install one in accordance with the governing electric code.
WARNING: Do not under any circumstances remove grounding conductor from the power plug.
WARNING: The power cable supplied with this equipment provides for this protection. Do not attempt to
defeat this protection by modifying the cable or by using ungrounded adapters or extension cables. The power cord
and plug must be intact and undamaged. To disconnect the equipment from the mains power; unplug the power
cord.
WARNING: Do not connect to an electrical outlet controlled by a wall switch or dimmer.
WARNING: If there is any doubt about the integrity of the protective ground conductor arrangement, do not
operate the monitor until the ac power source protective conductor is fully functional.
WARNING: For power interruptions exceeding 30 seconds, the monitor must be turned on manually by
pressing the Power On/Standby switch. When monitor power is restored, the monitor will return to manufacturer's
DEFAULT settings. (An option is available which will allow monitor to use the last used or STORED settings.)
WARNING: To avoid unacceptable RISK caused by power interruptions, connect the monitor to an
appropriate medical-grade uninterruptable power source (UPS).
WARNING: Do not place the monitor in any position that may cause it to fall on the patient. Do not lift the
monitor by the power supply cord or patient cable.
WARNING: Carefully route monitor cables (patient cables, power cords, etc.) to reduce the possibility of a
tripping hazard.
WARNING: Do not position the monitor in a way that would cause difficulty to the operator to disconnect it
from the power source.
WARNING: Electric shock hazard! Do not remove covers or panels. Refer service to trained and
qualified service personnel.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ivy Biomedical Systems 7600 and is the answer not in the manual?

Ivy Biomedical Systems 7600 Specifications

General IconGeneral
BrandIvy Biomedical Systems
Model7600
CategoryMedical Equipment
LanguageEnglish

Related product manuals