SAFETY
8 Model 7600/7800 Operation Manual
WARNING: The Model 7600/7800 should not be used adjacent to or stacked with other equipment.
However, if adjacent or stacked use is necessary, the Model 7600/7800 should be observed to verify normal
operation in the configuration in which it will used.
5.10 Accessories
WARNING: The use of accessories other than those specified below may result in increased emissions or
decreased immunity of the equipment.
Ivy P/N Description
590432 Low Noise, Four Lead ECG Patient Cable, 10 ft. Long, Colors: White, Green, Red, Black
590433 Set of Four Shielded Lead Wires, 24 Inches Long, Colors: White, Green, Red, Black
590435 Set of Four Radiotranslucent Lead Wires, 30 Inches Long, Colors: White, Green, Red, Black
590442 Set of Four Radiotranslucent Lead Wires, 36 Inches Long, Colors: White, Green, Red, Black
590446 Low Noise, Four Lead ECG Patient Cable, 10 ft. Long, IEC Colors: Red, Black, Green, Yellow
590447 Set of Four Shielded Lead Wires, 24 Inches Long, IEC Colors: Red, Black, Green, Yellow
590451 Set of Four Radiotranslucent Lead Wires, 30 Inches Long, IEC Colors: Red, Black, Green, Yellow
590452 Set of Four Radiotranslucent Lead Wires, 36 Inches Long, IEC Colors: Red, Black, Green, Yellow
590436 Radiotranslucent ECG electrodes, Box of 40 (10 pouches of 4 electrodes)
below. The customer or the user of the Model 7600/7800 should ensure that it is used in such an
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11 Radiated
Group 1
Class B
The Model 7600/7800 uses RF energy only for its
internal function. Therefore, their RF emissions are
very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11 Conducted
Class B
The Model 7600/7800 is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Class A

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ivy Biomedical Systems 7600 and is the answer not in the manual?

Ivy Biomedical Systems 7600 Specifications

General IconGeneral
BrandIvy Biomedical Systems
Model7600
CategoryMedical Equipment
LanguageEnglish

Related product manuals