SPECIFICATIONS
Model 7600/7800 Operation Manual
Resolution: Vertical - 200 dots/in.
Horizontal - 600 dots/in. at ≤25 mm/s
400 dots/in. at >25 mm/s
Frequency Response: >100 Hz at 50 mm/s
Data Rate: 500 samples
Synchronized Output (Trigger)
Test input signal at ECG leads: Conditions: ½ sine wave, 60ms width, 1mV amplitude, 1 pulse/second
Output Trigger Delay: < 2 ms
R to R Trigger Accuracy: ±75 µs typical @ 1 mV input
Pulse width: 1ms, 50ms, 100ms or 150ms (Configuration Dependent)
Pulse amplitude: 0V to +5V or -10V to +10V (Configuration Dependent)
Pulse amplitude polarity: Positive or Negative (Configuration Dependent)
Output Impedance: <100 Ω
Sensitivity and Threshold
Adjustment: Fully Automatic
Real Time Clock
Resolution: 1 minute
Display: 24 hours
Power Requirement: The real time clock keeps time whether the monitor has power or not.
The real time clock is powered by a dedicated lithium battery whose
life is a minimum 5 years at a temperature of 25
o
C
Note: The dedicated real time clock lithium battery is embedded in the
SNAPHAT package (not a bare battery) and therefore is considered
“contained in equipment”.
Operating Environment
Temperature Range: 5°C to 40°C
Relative Humidity: 0% to 90% non-condensing
Altitude: -100 meters to +3,600 meters
Atmospheric Pressure: 500-1060 mbar
Protection against ingress of fluids: IPX1 – Protection against vertically falling drops of water
Storage Environment
Temperature Range: -40°C to +70°C
Relative Humidity: 5% to 95%
Altitude: -100 meters to +14,000 meters
Power Requirements
Voltage Input: 100-120V~; 200-230V~
Line Frequency: 50/60 Hz
Fuse Rating and Type: T 0.5AL, 250V
Maximum ac Power
Consumption: 45 VA
Power Recovery: Automatic, if power is restored within 30 seconds