V03
JCAII 7
6) Press key to save the settings and return to weighing mode.
7-5 Calibration
= remote platform
= main scale
Note:
Please calibrate the main scale first.
Press key to switch the calibration for the main scale or remote platform before the zero
point calibration.
You can only do the calibration for the remote platform after you have set the max. capacity and
division of the remote platform.
Here take JCAII-6K as an example.
1. Press key and hold while turning on the scale. Do not release till
is displayed in the Unit Weight window.
2. Press key to start the zero point calibration and wait till is displayed in the Unit
Weight window. Input the first calibration weight value by the numeric key (6K model, 1/3 of full load
is 2kg). And then put the corresponding weights on.
3... Press key and wait till is displayed in the Unit Weight window. And then put
the corresponding weights on.
4. Press key and wait till is displayed in the Unit Weight window. And then put the
corresponding weights on.