Appendix
70
-427
Pressure sensor detects excessive
condensation pressure.
Check ambient temperature and
reduce if necessary.
Check condenser for soiling and clean
as necessary.
Switch off the unit at the mains switch,
wait 4 seconds and then switch the
unit on again.
For water-cooled units: Check cooling
water temperature and supply.
If the fault has not been remedied,
contact Technical Service.
-431 The maximum permissible current
consumption at the compressor has
been exceeded.
Check mains voltage for nominal
voltage.
If the fault has not been remedied,
contact Technical Service.
-1427 Pressure sensor detects excessive
compensation pressure.
Check ambient temperature and
reduce if necessary.
Check condenser for soiling and clean
as necessary.
Switch off the unit at the mains switch,
wait 4 seconds and then switch the
unit on again.
For water-cooled units: Check cooling
water temperature and supply.
If the fault has not been remedied,
contact Technical Service.
-1431 The minimum permissible current
consumption at the compressor has
been fallen short of.
Check mains voltage for nominal
voltage. The specified voltage
tolerance of the unit must not be
exceeded.
Check the mains cable of the
refrigeration unit for damage and
replace if necessary.
Check ambient temperature and
reduce if necessary.
Check CAN bus cable for damage and
replace as necessary.
If the fault has not been remedied,
contact Technical Service.