Appendix
71
-2426
Evaporation temperature has fallen
below the warning threshold.
Check pump setting
Check hose cross-section of connection
to application
Check that connection to application is
clear
Check suitability of bath fluid
11.3 Error messages in configuration process
If errors occur during a configuration process or during a firmware update, these
are shown on the display as error codes in ticker text.
Error code Description Solution
CFG error Error during configuration.
Repeat the process.
Replace USB stick if this occurs again.
If the fault has not been remedied,
contact Technical Service.
ProG error Error during firmware update.
Repeat the process.
Replace USB stick if this occurs again.
If the fault has not been remedied,
contact Technical Service.