Appendix
71
-2426
Evaporation temperature has fallen
below the warning threshold.
• Check pump setting
• Check hose cross-section of connection
to application
• Check that connection to application is
clear
•
Check suitability of bath fluid
11.3 Error messages in configuration process
If errors occur during a configuration process or during a firmware update, these
are shown on the display as error codes in ticker text.
Error code Description Solution
CFG error Error during configuration.
• Repeat the process.
• Replace USB stick if this occurs again.
• If the fault has not been remedied,
contact Technical Service.
ProG error Error during firmware update.
• Repeat the process.
• Replace USB stick if this occurs again.
• If the fault has not been remedied,
contact Technical Service.