Operating procedures
26
6. Operating procedures
6.1. Power connection
Caution:
Only connect the unit to a power socket with earthing contact (PE –
protective earth)!
The power supply plug serves as safe disconnecting device from the line
and must be always easily accessible.
Never operate equipment with damaged mains power cables.
Regularly check the mains power cables for material defects (e.g. for
cracks).
We disclaim all liability for damage caused by incorrect line voltages!
Check to make sure that the line voltage matches the supply
voltage specified on the identification plate.
Connect the circulator with mains power cable (20a) to the
mains outlet (21).
Connect the control cable (15a) between the connectors
(15, 15b).
Connect the refrigerated circulator with mains power cable
(20b) to the mains socket.
6.2. Switching on / Start – Stop
6.2.1. Switching on the circulator
Switching on:
Turn on the mains power switch (1).
L The unit performs a self-test. All segments of the 4-digit MULTI-
DISPLAY (LED) and all indicator lights will illuminate.
Then the software version (example: tt 2, V1.12, b004))
appears. The display „OFF“ or „R OFF“ indicates the unit is
ready to operate.
L The circulator enters the operating mode activated before
switching the circulator off:
keypad control mode (manual operation)
or
remote control mode (operation via personal computer).