MA
53
Command Parameter Response of circulator
in_sp_02 none Working temperature „t 3“
in_sp_03 none High temperature warning limit „t High“
in_sp_04 none Low temperature warning limit „t Low“
in_sp_07 none Pump pressure stage
in_par_01 none Te - Time constant of the external bath.
in_par_02 none Si - Internal slope
in_par_03 none Ti - Time constant of the internal bath.
in_par_06 none Xp control parameter of the internal controller.
in_par_07 none Tn control parameter of the internal controller.
in_par_08 none Tv control parameter of the internal controller.
in_mode_01 none Selected setpoint:
0 = Setpoint „t 1“
1 = Setpoint „t 2“
2 = Setpoint „t 3“
in_mode_05 none Circulator in Stop/Start condition:
0 = Stop
1 = Start
12.4. Status messages
Status messages Description
00 MANUAL STOP
Circulator in „OFF“ state.
01 MANUAL START
Circulator in keypad control mode.
02 REMOTE STOP
Circulator in „r OFF“ state.
03 REMOTE START
Circulator in remote control mode.
12.5. Error messages
ERROR MESSAGES Description
-01 LOW LEVEL ALARM Low liquid level alarm.
-02 REFRIGERATOR ALARM Control cable of the refrigerated circulator or
MVS solenoid valve controller short-circuited
or interrupted.
-03 EXCESS TEMPERATURE WARNING High temperature warning.
-04 LOW TEMPERATURE WARNING Low temperature warning.
-05 WORKING SENSOR ALARM Working temperature sensor short-circuited or
interrupted.