2.2. Technical specifications
F12-MA F25-MA
Working temperature range °C -20 ... 200 -28 ... 200
Temperature stability °C ±0,02 ±0,02
Cooling capacity
Medium ethanol
°C
kW
+20 0 -20
0.16 0.1 0.02
+20 0 -20
0.26 0.2 0.06
Refrigerant R134a R134a
Overall dimensions (WxDxH) cm 20x36x56 23x42x61
Bath opening (WxL) cm 13x15 12x14
Bath depth cm 13 14
Filling volume from ... to liters 3 ... 4,5 3 ... 4,5
Weight kg 23 31
Mains power connection
208V - 230V; 50Hz - 60Hz
V; Hz 187 - 253; 50 - 60
(perm. range)
187 - 253; 50 - 60
(perm. range)
Current draw (at 208V) A 11 12
Current draw (at 230V) A 11 12
Current draw (at 230V) CH A <9 + 1> <9 + 2>
Mains power connection
100V - 115V; 50Hz - 60Hz
V; Hz 90 - 127; 50 - 60
(perm. range)
90 - 127; 50 - 60
(perm. range)
Current draw (at 100V) A 15 13
Current draw (at 115V) A 12 13
F32-MA
Working temperature range °C -35 ... 200
Temperature stability °C ±0,02
Cooling capacity
Medium ethanol
°C
kW
+20 0 -20
0.45 0.39 0.15
Refrigerant R134a
Overall dimensions (WxDxH) cm 31x42x64
Bath opening (WxL) cm 18x12
Bath depth cm 15
Filling volume from ... to liters 5,5 ... 8
Weight kg 37
Mains power connection
208V - 230V; 50Hz - 60Hz
V; Hz 187 - 253; 50 - 60
(perm. range)
Current draw (at 208V) A 10
Current draw (at 230V) A 13
Mains power connection
100V - 115V; 50Hz - 60Hz
V; Hz 90 - 127; 50 - 60
(perm. range)
Current draw (at 100 V) A 14
Current draw (at 115 V) A 14
All measurements have been carried out at:
rated voltage and frequency ambient temperature: 20 °C
Technical changes without prior notification reserved.