Operator responsibility – Safety recommendations
10
Ultra-Low Refrigerated Circulator
FP55-SL
FPW55-SL
Working temperature range °C -60 ... 100
Special unit with extended
working temperature range
°C -60 ... 150
Temperature stability °C ±0.05
Cooling capacity
°C
+20 0 -20 -40 -60
Cooling compressor 1-stage
Refrigerant R404A
Cooling water flow rate at 20 °C l/h approx. 400
Overall dimensions (WxDxH) cm 85x76x116
59x76x116
Filling port dia. cm 7
Special unit with bath opening
cm 28x23
Bath depth cm 23
Filling volume liters 27
Weight kg 180 146
Ambient temperature °C 5 ... 40
Cooling water connection ------ G3/4“ male thread
Mains power connection V/ Hz 400 V/3PNPE/ 50 Hz 400 V/3PNPE/ 50 Hz
Current draw (at 400 V) A 20 20
Mains power connection V/ Hz 208 -230V/3PPE/ 60 Hz 208-230 V/3PPE/ 60 Hz
Current draw (at 208 V / 230 V) A 28 28
Current draw: Refer to notice page 15
All measurements have been carried out at: rated voltage and frequency ambient temperature: 20 °C
Technical changes without prior notification reserved.