EMC requirements
The device is an ISM device of group 1 per CISPR 11 (uses HF for internal purposes) and is
classified in class A (industrial and commercial sector).
Notice:
• Devices of class A are intended for the use in an industrial electromagnetic environment.
• When operating in other electromagnetic environments, their electromagnetic compatibility
may be impacted.
• This device is not intended for the use in living areas and cannot guarantee adequate
protection of the radio reception in such environments.
2.3. Mains power connections
Note concerning current consumption: Mains power connection 3PNPE and 3PPE
The indicated value of current consumption is the maximum useful current of the phase,
which also feeds the SL circulator with a 3 kW heater. If the ultra-low refrigerated circulator
is operated with the HST additional heater, the 6 kW heating capacity is divided in equal
shares (2x3 kW) on the two other phases..
Model Mains power connection 3PNPE
FP52-SL 400 V/3PNPE/50 Hz
FPW52-SL 400 V/3PNPE/50 Hz
FP55-SL 400 V/3PNPE/50 Hz
FPW55-SL 400 V/3PNPE/50 Hz
FP90-SL 400 V/3PNPE/50 Hz
FPW90-SL 400 V/3PNPE/50 Hz
FPW91-SL 400 V/3PNPE/50 Hz