SL
89
The OUT commands are valid only in remote control mode.
Examples:
When the RS485 interface is used, the three-digit instrument address
precedes each command.
(example: address Ad32 = A032)
Command to set the working temperature >Setpoint1< to 55.5 °C
OUT_SP_00
55.5
A032_OUT_SP_00
55.5
Command to retrieve the working temperature >Setpoint1<
IN_SP_00
A032_IN_SP_00
Response from the circulator:
55.5
LF
A032_55.5
LF
Notice: OUT commands
Temperature values can be displayed on the circulator in °C or °F.
The settings sent via the interface must also be made in the selected unit according
to this setting.
12.3. List of commands
OUT commands: Setting parameters or temperature values.
Command Parameter Response of circulator
OUT_MODE_01
0 Use working temperature >Setpoint1<
OUT_MODE_01
1
Use working temperature >Setpoint2<
OUT_MODE_01
2 Use working temperature >Setpoint3<
OUT_MODE_02
0 Selftuning “off”.
Temperature control using the stored parameters.
OUT_MODE_02
1 Selftuning “once”
Single selftuning of the controlled system after the next start.
OUT_MODE_02
2 Selftuning “always”
Continual selftuning of the controlled system whenever a new
setpoint is to be reached.
OUT_MODE_03
0 Set external programmer input to voltage.
Voltage 0 V ... 10 V
OUT_MODE_03
1 Set external programmer input to current.
Current 0 mA ... 20 mA
OUT_MODE_04
0 Temperature control of internal bath.
OUT_MODE_04
1 External control with Pt100 sensor.
OUT_MODE_05
0
Stop the unit = R OFF-.
OUT_MODE_05
1 Start the unit.