SL
29
5.3.1. Tubing
Recommended tubing:
CR
®
tubing 8 mm inner dia.
CR
®
tubing 12 mm inner dia.
Viton tubing 8 mm inner dia.
Viton tubing 12 mm inner dia.
Insulation for tubing 8 mm inner dia.
Insulation for tubing 12 mm inner dia.
8 930 210 1.0 m
8 930 211 1.5 m
Metal tubing, triple insulated, M16x1
8 930 221 1.0 m
8 930 222 1.5 m
Metal tubing, insulated, M16x1
At high working temperatures the tubing used for temperature application and cooling
water supply represents a danger source.
A damaged tubing line may cause hot bath fluid to be pumped out within a short time.
This may result in:
• Burning of skin
• Difficulties in breathing due to hot atmosphere
Safety recommendations
• Employ suitable connecting tubing.
• Make sure that the tubing is securely attached.
• Avoid sharp bends in the tubing, and maintain a sufficient distance from
surrounding walls.
• Regularly check the tubing for material defects (e.g. for cracks).
• Preventive maintenance: Replace the tubing from time to time.