Remote control
90
Command Parameter Response of circulator
0 Set the control dynamics - aperiodic
OUT_MODE_08
1 Set the control dynamics - standard
xxx.xx Set working temperature. “Setpoint 1”
xxx.xx Set working temperature. “Setpoint 2”
OUT_SP_02
xxx.xx Set working temperature. “Setpoint 3”
xxx.xx Set high temperature warning limit “OverTemp”
xxx.xx Set low temperature warning limit “SubTemp”
OUT_SP_06
xxx.xx Set manipulated variable for heater via serial interface
OUT_SP_07
x Set the pump pressure stage. (1 ... 4)
OUT_PAR_04
x.x CoSpeed for external control 0 ... 5.0.
OUT_PAR_06
xxx Xp control parameter of the internal controller. 0.1 ... 99.9
Tn control parameter of the internal controller. 3 ... 9999
xxx Tv control parameter of the internal controller. 0 ... 999
xxx Xp control parameter of the cascade controller. 0.1 ... 99.9
OUT_PAR_10
xxx Proportional share of the cascade controller. 1 ... 99.9
Tn control parameter of the cascade controller. 3 ... 9999
xxx Tv control parameter of the cascade controller. 0 ... 999
xxx Maximum internal temperature of the cascade controller.
OUT_PAR_14
xxx Minimum internal temperature of the cascade controller.
Band limit (upper) 0 ... 200 °C
xxx Band limit (lower) 0 ... 200 °C
OUT_HIL_00
-xxx Set the desired maximum cooling power (0% to 100%).
Note: Enter the value with a preceding negative sign!
This setting is meaningful only for FP cooling machines.
OUT_HIL_01
xxx Set the desired maximum heating power (10% to 100%).
IN commands: Asking for parameters or temperature values to be displayed.
Command Parameter Response of circulator
VERSION
None Version number of the software (V X.xx)
Status message, error message (see page 92)
none Current bath temperature.
IN_PV_01
none Heating power being used (%).