EasyManua.ls Logo

Julabo FP52-SL - Page 79

Julabo FP52-SL
103 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SL
79
After returning the LCD display to standard display by pressing the
key , the temperature value adjusted and set on the external
voltage or current source is displayed in line 1
(Example: Eprog 50.00 °C).
This EPROG input enables the use of different voltage and current
values as program parameters.
“L Value” - Setting the Low value: (See below
)
1) Adjust and set the lowest desired value on the voltage or
current source (Example A: 1 V).
Wait approximately 30 seconds.
2) Assign a lower temperature threshold value to this adjusted
voltage/current value by pressing the appropriate keys on the
keypad of the instrument (Example A: 20 °C) and set by
pressing
.
“H Value” - Setting the High value: (See below
)
1) Adjust and set the highest desired value on the voltage or
current source (Example A: 10 V).
Wait approximately 30 seconds.
2) Assign an upper temperature threshold value to this adjusted
voltage/current value by pressing the appropriate keys on the
keypad of the instrument
(Example A: 200 °C) and set by pressing .
Example B in the diagram illustrates that the end point values
are freely selectable (e.g., 8 mA and 16 mA).
Example out of diagram A:
Adjusting the voltage source for an output of 7.6 V!
Line 5 of the LCD DIALOG-DISPLAY shows the externally set setpoint
value. The instrument calculates this value from the slope of the two
specified end points (in example A, 7.6 V corresponds to an external
setpoint temperature of 152.0 °C ).
After returning the LCD display to standard display by pressing , this
value is displayed in line 1
(Example: EPROG 152.00 °C).
Notice:
If this adjustment is not correctly performed at two different points, the setpoint setting
will be incorrect.
Important:
The usable temperature range between > L Value < and > H Value < is limited to the
configured working temperature range of the circulator or unit combination. For the
working temperature range, see technical specifications.
20
40
60
80
100
200
300
250
150
°C
1 2 3 4 5 6 7 8 9 10 V
2 4 6
8
10 14 18 20 mA
A
B

Table of Contents

Related product manuals