5. OPERATION
GPI assignment
ACCENT has four GPI inputs for remote control. Following functions can be assigned to these
inputs (on if GPI input is switched to ground):
preset 1, preset 2, preset 3, preset 4
bypass on, compressor on, limiter on,
expander on, filter on, reset (clip display)
GPI is overwriting front panel settings.
Software version
display of software version D: dsp software C: controller software
5.10 RECALL AND STORAGE OF PRESETS (preset button)
All individual settings for the signal processing and the status of the unit can be stored as presets.
99 presets are storable into the unit.
Adjustments are made by pushing and turning SELECT knob (see 5.1).
PR LOAD RECALL SAVE
20 ________________
number of
selected preset
preset name max.
16 characters
LOAD is recalling the processing parameter only
(expander, filter, compressor, gain, limiter)
RECALL is recalling the processing parameter
and
the status of the unit
(expander, filter, compressor, gain, limiter
and configuration, input and output setup,
link, display, fader setup, GPI and tally
assignment)
Load/recall procedure
PRESET selection of PRESET menu
ush
SELECT selection of LOAD or RECALL mode
SELECT switches to preset number selection
turn
SELECT selection of preset number for loading or
recalling preset
push
turn
SELECT executes loading or recalling of preset,
exit preset menu
push
Push any other function button for leaving the preset menu without loading or recalling presets.
page 5-6 Operation manual accent2, chapter 5 -Operation-