Section 10
35
K-Link
10
SECTION 10 - K-LINK
K-Link soware can be used to communicate with a G210 over a TCP/IP network to retrieve,
display and save a log of measurements, warnings, and daily averages into a spreadsheet.
K-Link can also be configured to send email notifications when alarms are triggered.
Starting K-Link
To launch the K-Link soware, double click on the K-Link icon on the desktop or in the
start menu.
NOTE: In order modify settings in K-Link such as adding a new device to the Device
Connection list, or modifying email configuration settings, K-Link must be launched with
elevated administrative privileges.
To launch K-Link with elevated administrative
privileges, right-click on the K-Link icon, then click
“Run as administrator”. There may be a prompt
to enter alternative credentials. Consult the IT
Department for further details.
The K-Link loading screen is displayed for a few
seconds while the soware loads.
Device Connection
The device connection screen displays a list of previously saved devices to load.
A new device can be added to this list by entering its IP address, then aer performing a
successful connection test, pressing “Save”.
Aer loading a device, and performing a successful connectivity test, indicated by status “OK”,
logging can be started by pressing “Start”.

Table of Contents