Bladder Volume Tester User’s Manual V1.02
- 7 -
6.1 System placement
Please carefully read through and fully understand the safety cautions before moving and placing the
system.
1. Unpack the instrument case and check the goods for its completeness according to the packing list
furnished.
2. Place the instrument on a stable and leveled position.
3. Leave adequate space of 20 centimeters as minimum from rear, left and right side of the instrument.
6.2 Probe bracket installation
The probe bracket should be mounted on the both sides of the main unit.
6.3 Ultrasonic probe installation
6.3.1 Ultrasonic probe connection
Warning:
Before connecting or using the probe, make sure that the probe, connecting
cable and connector are in normal condition (free of cracks or drop). Use of defective
Attention:
Adequate space from rear, left and right side of the machine shall be reserved,
or the machine may malfunction under excessive heat inside the enclosure.
Attention:
1. Usually the probe should be placed within the probe bracket, not on the desktop or other
support to avoid the drop.
2. Probe is a critical, precision part, do not stress, impact, or fall it. Do not pull or wring
wound probe cable.
3. Turn off the ultrasonic system before disconnecting the probe. Disconnecting the probe
with system power on may damage the system or probe.
4. Before disconnecting the probe, place the probe on the probe bracket in order to prevent
the probe may not be damaged by unexpected fall.
5. When the instrument is power on but not in use, please make it in a frozen state, in order to
increase of service life of probe.
6. Repeat available machine time should be more than 5 minutes to avoid turn on/off power
supply in short time.
Attention:
The red mark of connector should be aligned with the red mark on the socket
when inserting the probe.
Warning:
1. Do not use the probe not provided by our company, otherwise the equipment and the probe
will cause damage, and may cause fire in extreme cases.
2. Check the ultrasonic probe and connecting cable after diagnostic operation. Use of
defective probe may cause electric shock.
3. Do not strike the probe; using the damaged probe may cause electric shock to the patient.
4. Unauthorized disassembly of the probe shall be prohibited as it may cause electric shock.
Danger:
Use together with flammable anaesthetic, it may result in explosion.
Attention:
Probe is highly sensitive to shake, be used with caution. About probe’