Bladder Volume Tester User’s Manual V1.02
- 23 -
Fig. Probe regular disinfection
3. Check probe after cleaning and disinfection
(1) Check the probe enclosure and its cable to ensure they are free of abnormity (such as scuffing,
cracks or drop-off);
(2) The sound window of probe is thin; ensure that there are no any abnormity on the sound
window, such as scuffing, cracks, peeling, and bulge.
12.2.2 Clean the probe socket
1. Clean the probe socket with soft, dry cloth;
2. In case of die-hard blots, clean with soft cloth dipped in moderate detergent and then air-dry it.
12.2.3 Clean the touch screen
Clean the liquid crystal display with dry, soft flax or anti-static LED clean cloth.
12.2.4 Clean the foot switch
1. Use the soft dry cloth to wipe the foot switch;
2. If it is difficult to wipe away the blemish, clean with soft cloth dipped in moderate detergent and
then air-dry it.
12.2.5 Clean the shell and probe bracket
Clean the instrument surface with soft, dry cloth or with soft cloth dipped in moderate water cleaning
media to remove the blots, and then dry the instrument with soft, dry cloth or with air.
12.2.6 Clean the thermal printer
When appeared unclear handwriting after the thermal printer used for some time, according to the
following steps to clean the print-head.
1. Remove the paper from the print-head, raised its head to expose the print-head piece;
2. Clean the dirt on the heating element surface of the print-head piece with a clean cotton swab dipped
in a little alcohol. Remember not to use hard objects to clean such as sandpaper, blade or screwdriver;
otherwise it would permanently damage the print-head piece.
3. Clean the print-head piece and then air-dried.
Attention:
Do not clean the screen with hydrocarbon detergent such as alcohol or OA
equipment cleaning media.
Attention:
Prohibit using sharp objects to touch the touch screen, and prohibit
pressing or squeezing against the touch screen.
Attention
:
1. It is a normal phenomenon that color of the acoustic lens may change and color of the
probe label may fade away.
2.
The regular disinfection times should be minimized as it may lead to degrade of the probe
safety and performance.
Prohibit immersion