Bladder Volume Tester Users Manual V1.02
- 30 -
Table 2 - Guidance and manufacturer’s declaration—electromagnetic immunity
The equipment is intended for use in the electromagnetic environment specified below. The
customer or the user should assure that they are used in such an environment.
Immunity test IEC 60601 test
level
Compliance level Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for signal
lines
± 2 kV for power
supply lines
± 1 kV for signal
lines
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode
± 2 kV common
mode
± 1 kV differential
mode
± 2 kV common
mode
Mains power quality should
be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
5U
T
(95dip in U
T
)
for 0.5 cycle
40U
T
(60dip in U
T
for 5 cycles
70U
T
(30dip in U
T
for 25 cycles
5U
T
(95dip in U
T
)
for 5 sec
5U
T
(95dip in U
T
for 0.5 cycle
40U
T
(60dip in U
T
for 5 cycles
70U
T
(30dip in U
T
for 25 cycles
5U
T
(95dip in U
T
)
for 5 sec
Mains power quality should
be that of a typical
commercial or hospital
environment.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3A/m 3A/m Mains power quality should
be that of a typical
commercial or hospital
environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.