EasyManua.ls Logo

Kantronics KPC-3 Plus - Calibration;Equalization

Default Icon
236 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
216
5. Observe on the computer display (your terminal program must be set at 1200 baud):
CHECKSUM OK
RAM OK
100,000 BYTES
NO CLOCK
REPLACE TEST JUMPER
6. Turn power off.
7. Return jumper J11 to the normal position (one pin only).
8. Reassemble the KPC-3 Plus and return to operation.
Note: After a hard reset, all parameters are at factory default values and any mailbox
messages are lost.
Calibration/Equalization
The CALIBRAT command is used to assist in adjusting the data transmitter drive level
and also in adjusting the receive equalization of your KPC-3 Plus. The procedures listed
below are a guideline for using the calibration command to adjust for optimum
performance.
When you issue the CALibrate command, the following options will appear on your
screen:
CALIBRATE MODE:
M send mark
R receive
S send space
T send square wave
- or + adjust XMITLVL while transmitting
X exit
Pressing M will transmit a constant MARK (signal) until another key is pressed.
Pressing R will measure a received square wave.
Pressing S will transmit a constant SPACE until another key is pressed.
Pressing T will transmit a square wave (space/mark) until another key is pressed.
Pressing X will exit the calibrate mode and reset the KPC-3 Plus.
Drive level.
The drive level is set by a digital pot, providing a range of 1 mV to 4 V p-p. Default drive
level is about 50 mV, corresponding to a XMITLVL command setting of 100. After
entering the calibrate mode (CAL) you should use the T option to initiate the
transmission of a square wave and then adjust the AFSK drive level by pressing the ‗-ā€˜

Table of Contents

Related product manuals