EasyManuals Logo
Home>Keithley>Multimeter>2000

Keithley 2000 Quick Start Guide

Keithley 2000
51 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
1-14 Return to Section Topics 2000-903-01 Rev. C / October 2007
Section 1: Introduction Model 2000 Digital Multimeter Quick Start Guide
Figure 1-3
Operation event status
Figure 1-4
Measurement event status
OR
Idle = Idle state of the 2000
Trig = Triggering
Meas = Measuring
& = Logical AND
OR = Logical OR
(B15 - B11)
Idle
(B10) (B9) (B8) (B7) (B6)
Trig
(B5)
Meas
(B4) (B3) (B2) (B1) (B0)
Operation Event
Register
(B15 - B11)
Idle
(B10) (B9) (B8) (B7) (B6)
Trig
(B5)
Meas
(B4) (B3) (B2) (B1) (B0)
Operation Event
Enable Register
&
To Operation
Summary Bit
(OSB) of Status
Byte Register.
(See Figure 6).
&
&
(B15 - B11)
Idle
(B10) (B9) (B8) (B7) (B6)
Trig
(B5)
Meas
(B4) (B3) (B2) (B1) (B0)
Operation
Condition Register
OR
BFL = Buffer Full
BHF = Buffer Half Full
BAV = Buffer Available
& = Logical AND
OR = Logical OR
(B15 - B12) (B10) (B9)
BHF
(B8)
BAV
(B7) (B6)
RAV
(B5) (B4) (B3)
(B2)
LL
(B1) (B0)
Measurement Event
Register
(B15 - B12)
Measurement Event
Enable
Register
To Measurement
Summary Bit
(MSB) of Status
Byte Register.
(See Figure 6).
&
&
(B15 - B12) (B10) (B9) (B8) (B7) (B6)
RAV
(B5) (B4) (B3)
(B2)
LL
(B1) (B0)
Measurement
Condition Register
BFL
HL
ROF
BFL
ROF
&
&
&
BHF BAV
HL
&
&
(B10) (B9)
BHF
(B8)
BAV
(B7) (B6)
RAV
(B5)
(B4) (B3)
(B2)
LL
(B1) (B0)
BFL ROFHL
RAV = Reading Available
HL = High Limit
LL = Low Limit
ROF = Reading Overflow
(B11)
(B11)
(B11)

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
BrandKeithley
Model2000
CategoryMultimeter
LanguageEnglish

Related product manuals