EasyManua.ls Logo

Keithley 2000 - Power-up

Keithley 2000
51 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2000-903-01 Rev. C / October 2007 Return to Section Topics 1-15
Model 2000 Digital Multimeter Quick Start Guide Section 1: Introduction
Figure 1-5
Questionable event status
Figure 1-6
Status byte and service request (SRQ)
(B13 - B9)(B14) (B8)(B15)
OR
Questionable
Condition Register
Questionable Event
Enable Register
Warn = Command Warning
Cal = Calibration Summary
Temp = Temperature Summary
& = Logical AND
OR = Logical OR
&
&
&
0
(B13 - B9)(B14)
(B8)
(B15)
0
Warn Cal
Questionable Event
Register
(B13 - B9)(B14) (B8)(B15)
To Questionable
Summary Bit (QSB)
of Status
Byte Register
(See Figure 6).
(B7 - B5)
(B7 - B5)
(B7 - B5)
Warn Cal
0
Warn Cal
(B4)
Temp
(B4)
Temp
(B3 - B0)
(B3 - B0)
(B3 - B0)(B4)
Temp
&
Status Summary Messages
* STB?
Serial Poll
OSB
(B7)
RQS
(B6)
MSS
ESB
(B5)
MAV
(B4)
QSB
(B3)
EAV
(B2)
(B1) (B0)
OR
* SRE
* SRE?
Status Byte
Register
Service
Request
Enable
Register
OSB = Operation Summary Bit
MSS = Master Summary Status
RQS = Request for Service
ESB = Event Summary Bit
MAV = Message Available
QSB = Questionable Summary Bit
EAV = Error Available
MSB = Measurement Summary Bit
& = Logical AND
OR = Logical OR
OSB
(B7) (B6)
ESB
(B5)
MAV
(B4)
QSB
(B3)
EAV
(B2)
(B1) (B0)
&
&
&
&
&
MSB
MSB
&
Service
Request
Generation
Read by Serial Poll
Read by *STB?

Other manuals for Keithley 2000

Related product manuals