2000-903-01 Rev. C / October 2007 Return to Section Topics 1-27
Model 2000 Digital Multimeter Quick Start Guide Section 1: Introduction
:PERiod
:APERture
:APERture?
:THReshold
:VOLTage
:RANGe <n>
:RANGe?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
:DIGits <n>
:DIGits?
Path to configure period.
Sets gate time for period
measurements (0.01-1.0s).
Query period gate time.
Path to select the threshold
voltage range:
Select threshold range (0 to
1010).
Query threshold range.
Specify reference (0 to 1).
Enable or disable reference.
Query state of reference.
Use input signal as
reference.
Query reference value.
Specify measurement
resolution (4 to 7).
Query resolution.
:DIODe
:CURRent
:RANGe
[:UPPer] <NRf>
[:UPPer]?
Paths to configure diode test:
Path to select range.
Select range (0 to 1e-3)
Query range.
:CONTinuity
:THReshold <NRf>
:THReshold?
Path to configure continuity
test:
Set threshold resistance (1 to
1000).
Query threshold resistance.
Table 1-10
SENSe command summary (cont.)
Command Description