3: Performance verification Model 2110 5½ Digit Multimeter
3-2 2110-905-01 Rev. C / August 2013
AC voltage accuracy verification
AC VOLTAGE ACCURACY VERIFICATION
Range Test Point Accuracy Limits
Upper Limit Lower Limit
100 mV 100.0000 mV @ 10 Hz 100.170 mV 99.830 mV
100.0000 mV @ 30 kHz 100.300 mV 99.700 mV
10 V 10.00000 V @ 10 Hz 10.02 V 9.98 V
10.00000 V @ 30 kHz 10.03 V 9.97 V
100.0000 V @ 200 kHz 105.00 V 95.00 V
Specifications valid for 2-hour warm-up at 5-1/2 digits:
a. Slow ac filter (3 Hz bandwidth).
b. Pure sine wave input greater than 5% of range.
c. 750 V range limited to 100 kHz.