Section
1: Calibrating and adjusting the Model 2110 Model 2110 5½ Digit Multimeter
Calibration Manual
1-2 2110-905-01 Rev. C / August 2013
Figure 1: Model 2110 KI-TOOL software
5. The procedure consists of connecting the Fluke 5700A/5725A and 5520A/5522A output to the
Model 2110 front panel input, setting up appropriate ranges, and performing calibration by
sending specific SCPI commands. These commands are entered into the Send String field
followed by clicking the Write button. To access the command control window, click Tool >
Command Control.
The figure below shows the Command Control window. Ensure that the correct model number is
selected before sending any SCPI commands.