2-2 Calibration Model 2500 Service Manual
 
Introduction
 
Use the procedures in this section to calibrate the Model 2500 Dual Photodiode Meter. 
These procedures require accurate test equipment to supply precise currents and measure 
accurate DC voltages. Calibration can be performed either from the front panel or by send-
ing SCPI calibration commands over the IEEE-488 bus or RS-232 port with the aid of a 
computer.
 
WARNING
 
The information in this section is intended for qualified service person-
nel only. Do not attempt these procedures unless you are qualified to do 
so. Some of these procedures may expose you to hazardous voltages.
 
Environmental conditions
 
Temperature and relative humidity
 
Conduct the calibration procedures at an ambient temperature of 65-82°F (18-28°C) with 
relative humidity of less than 70% unless otherwise noted.
 
Warm-up period
 
Allow the Model 2500 to warm up for at least one hour before performing calibration.
If the instrument has been subjected to temperature extremes (those outside the ranges 
stated above), allow additional time for the instrument’s internal temperature to stabilize. 
Typically, allow one extra hour to stabilize a unit that is 50°F (10°C) outside the specified 
temperature range. 
Allow the test equipment to warm up for the minimum time specified by the manufacturer.
 
Line power
 
The Model 2500 requires a line voltage of 100V / 120V / 200V / 240V (depending on line 
voltage setting) at line frequency of 50 or 60Hz. The instrument must be calibrated while 
operating from a line voltage within this range. Make sure the line voltage setting seen 
through the small window in the rear panel power module is at the required setting. If not, 
change the setting as covered in Section 3.