B-6 Calibration Reference Model 2500 Service Manual
Current measurement commands
:SENSe
(:CALibration:PROTected:SENSe[1])
(:CALibration:PROTected:SENSe2)
Purpose To calibrate the active channel 1 or channel 2 current range.
Format :cal:prot:sens1 <Cal_current>
:cal:prot:sens2 <Cal_current>
Parameters <Cal_current> = Within ±10% of positive full-range value
0 ±0.1% of full-range value
Within ±10% of negative full-range value
Description The :CAL:PROT:SENS1 and :CAL:PROT:SENS2 commands calibrate 
the active range of the channel 1 or channel 2 current measurement 
respectively. During the calibration process, this command is sent three 
times for each range, once each with parameters of approximately 
positive full range, 0, and negative full range. The appropriate 
calibration current must be applied to the channel 1 or channel 2 input 
terminals.
Example :CAL:PROT:SENS1 1.9e-3 Calibrate channel 1 2mA range.
:DATA
(:CALibration:PROTected:SENSe[1]:DATA)
(:CALibration:PROTected:SENSe[1]:DATA?)
(:CALibration:PROTected:SENSe2:DATA)
(:CALibration:PROTected:SENSe2:DATA?)
Purpose To set/query channel 1 or channel 2 measurement calibration constants.
Format :cal:prot:sens1:data <Cal_constants>
:cal:prot:sens2:data <Cal_constants>
Parameters <Cal_constants> = ASCII floating point format, delimited by commas
Query :cal:prot:sens1:data?
:cal:prot:sens2:data?
Description The :CAL:PROT:SENS1:DATA and :CAL:PROT:SENS2:DATA 
commands allow you to send calibration constants directly to the unit, 
while the :CAL:PROT:SENS1:DATA? and :CAL:PROT:SENS2:DATA? 
queries request calibration constants. These commands allow you to 
alter calibration directly without having to perform the entire calibration 
procedure.
Example :CAL:PROT:SENS1:DATA? Query channel 1 constants.