EasyManuals Logo

Keithley 2500 Service Manual

Keithley 2500
121 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #102 background imageLoading...
Page #102 background image
Model 2500 Service Manual Calibration Reference B-5
:DATE
(:CALibration:PROTected:DATE)
Purpose To program the calibration date.
Format :cal:prot:date <year>, <month>, <day>
Parameters <year> = 2000 to 2099
<month> = 1 to 12
<day> = 1 to 31
Query :cal:prot:date?
Response <year>, <month>, <day>
Description The :DATE command allows you to store the calibration date in instru-
ment EEROM for future reference. You can read back the date from the
instrument by using the :DATE? query, or by using the front panel CAL
menu.
Note The year, month, and day parameters must be delimited by commas.
Example :CAL:PROT:DATE 2000,11,20 Send cal date (11/20/2000).
:NDUE
(:CALibration:PROTected:NDUE)
Purpose To send the next calibration due date to the instrument.
Format :cal:prot:ndue <year>, <month>, <day>
Parameters <year> = 2000 to 2099
<month> = 1 to 12
<day> = 1 to 31
Query :cal:prot:ndue?
Response <year>, <month>, <day>
Description The :NDUE command allows you to store the date when calibration is
next due in instrument memory. You can read back the next due date by
using the :NDUE? query, or by using the front panel CAL menu.
Note The next due date parameters must be delimited by commas.
Example :CAL:PROT:NDUE 2001,11,20 Send due date (11/20/2001).

Table of Contents

Other manuals for Keithley 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2500 and is the answer not in the manual?

Keithley 2500 Specifications

General IconGeneral
BrandKeithley
Model2500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals