EasyManua.ls Logo

Keithley 2500 - Example Limits Calculation; Restoring Factory Defaults

Keithley 2500
121 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 2500 Service Manual Performance Verification 1-5
Example limits calculation
As an example of how verification limits are calculated, assume you are testing the 2mA
range with a 1.9mA input current. Using the Model 2500 one-year accuracy specification
of ±(0.1% of reading + 50nA offset), the calculated reading limits are:
Limits = 1.9mA ± [(1.9mA
×
0.1%) + 2µA]
Limits = 1.9mA ± (1.9
µ
A + 2µA)
Limits = 1.9mA ± 3.9
µ
A
Limits = 1.8961mA to 1.9039mA
Restoring factory defaults
Before performing the verification procedures, restore the instrument to its factory front
panel (BENCH) defaults as follows:
1. Press the MENU key. The instrument will display the following prompt:
MAIN MENU
SAVESETUP COMMUNICATION CAL
4
2. Select SAVESETUP and then press ENTER. The unit then displays:
SAVESETUP MENU
SAVE RESTORE POWERON RESET
3. Select RESET and then press ENTER. The unit displays:
RESET ORIGINAL DFLTS
BENCH GPIB
4. Select BENCH and then press ENTER to restore BENCH defaults.

Table of Contents

Other manuals for Keithley 2500

Related product manuals