EasyManuals Logo

Keithley 2500 Service Manual

Keithley 2500
121 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #30 background imageLoading...
Page #30 background image
Model 2500 Service Manual Calibration 2-3
Calibration considerations
When performing the calibration procedures:
Make sure the test equipment is properly warmed up and connected to the Model
2500 input or output terminals as required.
Allow signals to settle before calibrating each point.
Do not connect test equipment to the Model 2500 through a scanner or other
switching equipment.
If an error occurs during calibration, the Model 2500 will generate an appropriate
error message. See Appendix B for more information.
WARNING
The maximum common-mode voltage (voltage between LO and
chassis ground) is 200V DC. Exceeding this value may cause a shock
hazard.
CAUTION
Maximum signal on OUTPUT connectors is 100V @ 20mA DC.
Maximum signal on INPUT connectors is 20mA. Exceeding these
values may result in damage to the instrument.
Calibration cycle
Perform calibration at least once a year to ensure the unit meets or exceeds its
specifications.
Recommended calibration equipment
Table 2-1 lists the recommended equipment for the calibration procedures. You can use
alternate equipment as long as that equipment has specifications at least as good as those
listed in the table. For optimum calibration accuracy, test equipment specifications should
be at least four times better than corresponding Model 2500 specifications.

Table of Contents

Other manuals for Keithley 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2500 and is the answer not in the manual?

Keithley 2500 Specifications

General IconGeneral
BrandKeithley
Model2500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals