Model 2500 Service Manual Calibration 2-7
Resetting the calibration password
If you lose the calibration password, you can unlock calibration by shorting together the 
CAL pads which are located on the display board. Doing so will also reset the password to 
the factory default (002500, front panel; KI002500, remote).
See Section 5 for details on disassembling the unit to access the CAL pads. Refer to the 
display board component layout drawing at the end of Section 6 for the location of the 
CAL pads.
Viewing calibration dates and calibration count
When calibration is locked, only the UNLOCK and VIEW-DATES selections will be 
accessible in the calibration menu. To view calibration dates and calibration count at any 
time:
1. From normal display, press MENU, select CAL, and then press ENTER. The unit 
will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES 
2. Select VIEW-DATES and then press ENTER. The Model 2500 will display the 
next and last calibration dates and the calibration count as in the following 
example:
NEXT CAL: 04/15/2001
Last cal: 04/15/2000  Count: 0001