7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-362 2600BS-901-01 Rev. B / May 2013
tsplink.trigger[N].stimulus
This attribute specifies the event that causes the synchronization line to assert a trigger. This attribute is not
available on the Models 2604B/2614B/2634B.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Instrument reset
Recall setup
TSP-Link trigger N reset
Not saved
0
Usage
eventID = tsplink.trigger[N].stimulus
tsplink.trigger[N].stimulus = eventID
eventID
The event identifier for the triggering event
N
The trigger line (1 to 3)
Details
To disable automatic trigger assertion on the synchronization line, set this attribute to zero (0).
Do not use this attribute when triggering under script control. Use tsplink.trigger[N].assert() instead.