Series 2600B
System SourceMeter® Instrument Reference Manual Appendix E:
Status model
2600BS-901-01 Rev. B / May 2013 E-23
Questionable Status Registers
This register set feeds to bit B3 (QSB) of the Status Byte. The bits used in the Questionable Status
Register set are described as follows:
• Bit B8, Calibration (CAL): Set bit indicates that calibration is questionable.
• Bit B9, Unstable Output (UO): Set bit indicates that an unstable output condition was detected.
• Bit B12, Over Temperature (OTEMP): Set bit indicates that an over temperature condition was
detected.
• Bit B13, Instrument Summary (INST): Set bit indicates that a bit in the Questionable Status
Instrument Summary Register is set.
For more information on the Questionable Status Register, refer to Status register set contents (on
page E-1) and the figures in this appendix.
Questionable Status Registers
As shown in the status model's Operation event, I/O, and TSP-Link registers (on page E-12), there
are seven register sets associated with Questionable Status. Commands are summarized in the
Status byte and service request (SRQ) (on page
E-15) topic. Note that bits can also be set by using
numeric parameter values. For details, see Programming enable and transition registers (on page E-
14).
For example, either of the following commands will set the CAL enable bit (B8):
status.questionable.enable = status.questionable.CAL
status.questionable.enable = 256
When reading a register, a numeric value is returned. The binary equivalent of this value indicates
which bits in the register are set. For details, see Reading registers (on page E-15
). For example, the
following command will read the Questionable Status Enable Register:
print(status.questionable.enable)
For more information about the Questionable Status Registers, refer to Status register set contents
(on page E-1) and the figures in this appendix.

Table of Contents

Other manuals for Keithley 2614B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2614B and is the answer not in the manual?

Keithley 2614B Specifications

General IconGeneral
BrandKeithley
Model2614B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals