EasyManuals Logo

Keithley 2614B Reference Manual

Keithley 2614B
864 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #777 background imageLoading...
Page #777 background image
Series 2600B
System SourceMeter® Instrument Reference Manual Appendix B:
Calibration
2600BS-901-01 Rev. B / May 2013 B-35
F. Select the DMM DC volts function.
G. Repeat the 22 steps of C for the 3 A and 10 A ranges (Model 2601B/2602B/2604B) or 1.5 A and
10 A ranges (Model 2611B/2612B/2614B/2634B/2635B/2636B). Compute the current reading
from the DMM voltage reading and characterized 0.5 Ω resistance value: I = V/R.
H. Select automatic polarity mode:
smua.cal.polarity = smua.CAL_AUTO
Models 2634B, 2635B, and 2636B:
1. Connect the Series 2600B to the digital multimeter. Use the figure titled "Connections for current
calibration (1.5 A through 10 A ranges)" as a guideline, but replace the 0.5 Ω resistor with the 1
GΩ resistor.
2. Select the multimeter DC current function.
3. Calibrate the low current ranges (100 pA, 1 nA, 10 nA, 100 nA, see Note) using a suitably
guarded and characterized 1 GΩ resistance standard, such as the Keithley Instruments Model
2600-STD-RES (see Recommended calibration adjustment equipment (on page B-20
)).
Step-by-step procedures, connection diagrams, and a factory script for calibrating the low current
ranges are included with the Model 2600-STD-RES. The general process entails forcing a
characterized voltage across the 1 GΩ resistor and comparing the Model 2634B/2635B/2636B
measured results against the standard resistance and voltage derived current.
The 2601B/2602B/2604B/2611B/2612B/2614B could be calibrated with this method for the 100 nA
setting if desired.
4. Characterize the appropriate +/- V source values with the Digital Multimeter according to the
Model 2634B/2635B/2636B calibration Step sequence (on page B-22
) .
5. Characterize the desired Model 2634B/2635B/2636B current ranges.
a. Connect the guarded resistance standard.
b. Source the appropriate voltage for +/- full-scale reading.
c. Wait 30 seconds for stable measurement.
d. Capture the Model 2634B/2635B/2636B reported current measurement.
e. Initiate HI-Z mode to open the resistor standard (source zero current) and the characterize offset.
f. Repeat the above steps for each low current range.
Settings of Model 2634B/2635B/2636B characterization of voltage source
Low current range Voltage source Compliance
100 pA*
±90.000 mV
1.5 A
1 nA
±0.90000 V
1.5 A
10 nA
±9.0000 V
1.5 A
100 nA
±90.000 V
100 mA
* Models 2635B and 2636B only.

Table of Contents

Other manuals for Keithley 2614B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2614B and is the answer not in the manual?

Keithley 2614B Specifications

General IconGeneral
BrandKeithley
Model2614B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals