10 7½ Digit Graphical Sampling Multimeter Calibration Manual Section 2:
Performance verification
DMM7510-905-01 Rev. A / April 2015 2-7
Monitoring internal temperature
You can monitor the temperature difference between the actual internal temperature and the
temperature when auto calibration ran through the panel or by using remote commands. With remote
commands, you can also check the present internal temperature and the internal temperature when
auto calibration was last run. Temperature is returned in Celsius (°C).
The internal temperature is not updated on the Calibration screen until the warmup period is
complete. The remote commands always return the present temperature.
From the front panel:
1. Press the MENU key.
2. Under System, select Calibration.
3. The Temperature Difference is displayed.
Using SCPI commands:
For the present internal temperature, send:
:SYSTem:TEMPerature:INTernal?
For the temperature difference, send:
:ACAL:LASTrun:TEMPerature:DIFFerence?
For the temperature when auto calibration was last run, send:
:ACAL:LASTrun:TEMPerature:INTernal?
Using TSP commands:
For the present internal temperature, send:
print(localnode.internaltemp)
For the temperature difference, send:
print(acal.lastrun.tempdiff)
For the temperature when auto calibration was last run, send:
print(acal.lastrun.internaltemp)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
Resolution7.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
Resistance Range10 Ω to 1 GΩ
Sampling Rate1 MS/s
Memory27.5 million readings
Display5 inch capacitive touchscreen
AC Voltage Accuracy±0.06% of reading
AC Current Range10 A
ConnectivityUSB, LAN, GPIB

Related product manuals